Background Worldwide more than 3 million people have cardiac pacemakers. Over 600,000 new pacemakers are implanted each year, with most of these devices in patients over the age of 60. Although pacemakers were once primarily used to treat bradyarrhythmias (e.g. heart block), more recently, patients with subvalvular stenosis, and atrial fibrillation may qualify for pacemakers. Additionally, patients with congestive heart failure may receive biventricular pacemakers or cardiac resynchronization therapy devices (CRT) to improve symptoms. This Fast Fact discusses management of cardiac pacemakers at life’s end. Fast Fact #112 discusses implantable cardioverter-defibrillators.
Pacemaker Function at Time of Death Patients and their families often make assumptions that pacemakers prolong the dying process and thus prolong suffering. However, a pacemaker is not a resuscitative device. In general, pacemakers do not keep dying patients alive, as terminal events are often due to sepsis, hemorrhage, pulmonary emboli, or arrhythmias from metabolic abnormalities associated with end-stage cancer, liver, or renal failure. At the time of death, the myocardium is usually too sick to respond to the pacemaker generated signals.
When is Pacemaker Deactivation Indicated? In patients with irreversible cognitive failure, where continued pacemaker activity is not meeting the goals of care, it may be appropriate to discuss the option of deactivation. In most other situations, deactivation is not indicated since the result is likely to be a symptomatic bradycardia, producing signs and symptoms of worsening heart failure (fatigue, dizziness, dyspnea). In contrast to popular belief, it is rare that disabling the pacemaker will result in a swift and painless death as few patients are 100% pacemaker dependant, particularly during the period of imminent death (Fast Fact # 3), where tachycardia is the most common rhythm. When questions arise concerning dependency on the pacemaker, consult the cardiology/pacemaker service.
CRT Devices These devices are used most often for patients with heart failure and, more rarely, in select patients with atrial fibrillation. CRT devices deliver biventricular pacing to improve the efficiency and functionality of the right and left heart ventricles. There are two main types: a) CRTs without capability for defibrillation (CRT-P) or b) CRTs with added capability for defibrillation (CRT-D). Most in the United States are of the CRT-D type. Below are relevant considerations:
·A CRT-P cannot be upgraded to CRT-D without lead replacement; however, a CRT-D may be downgraded to CRT-P in one of three ways: simple reprogramming; generator replacement; or placing a magnet over the generator (similar to an ICD).
·Indications for downgrade of CRT-D to CRT-P are evolving but often depend upon goals of care and prognosis. Downgrading does not affect the symptomatic benefits of pacing.
·While a trained radiologist or cardiologist may be able to discern the thicker RV lead and the visible coil of a CRT-D device on a chest x-ray, clinical distinction between CRT-D and CRT-P is best achieved by interrogating the device. This is usually performed by a technician from the manufacturing company or an electrophysiology clinician. Reference #7 provides an algorithm for identification of cardiac rhythm devices, which can facilitate this practical process.
Ethical/Legal issues A patient’s/surrogate’s right to request withdrawal of life sustaining medical interventions, including pacemakers, is both legal and ethical. Withdrawal of a life sustaining medical intervention with the informed consent of a patient or legal surrogate is not physician-assisted suicide or euthanasia. While there may be more agreement about the deactivation of implantable cardioverter-defibrillators than of pacemakers amongst practicing clinicians, the Heart Rhythm Sociecty issued a consensus statement in collaboration with many professional groups (including the American College of Cardiology, the American Academy of Hospice and Palliative Medicine, and the American Heart Association) which effectively erased any ethical distinction between types of implanted devices being deactivated and endorsed a patient’s right to have one’s pacemaker deactivated.
Summary Initiate a discussion about pacemaker deactivation only if there is potential for patient benefit; consider the potential negative effects of deactivation before disabling the pacemaker. For patients who wish to deactivate the ICD function of a CRT-D devices, this can often be done via simple reprogramming of the CRT device.
References
- Braun TC, Hagen NA, Hatfield RE, Wyse DG. Cardiac pacemakers and implantable defibrillators in terminal care. J Pain Symptom Manage. 1999; 18:126-131.
- Kolarik RC, Arnold RM, Fischer GS, Tulsky, JA. Objectives for advance care planning. J Palliative Med. 2002; 5(5):697-704.
- Mueller PS, Hook CC, Hayes, DL. Ethical analysis of withdrawal of pacemaker or implantable cardioverter-defibrillator support at the end of life. Mayo Clin Proc. 2003; 78:959-963.
- Wood MA, Ellenbogen KA. Cardiac pacemakers from the patient’s perspective. Circulation. 2002; 105:1022-1024.
- R. Lampert, D.L. Hayes, G.J. Annas, et al. HRS expert consensus statement on the management of cardiovascular implantable electronic devices (CIEDs) in patients nearing end of life or requesting withdrawal of therapy. Heart Rhythm 2010 (7), 1008-1026.
- Lindvall C, Chatterjee NA, Chang Y et al. National Trends in the Use of Cardiac Resynchronization Therapy With or Without Implantable Cardioverter-Defibrillator. Circulation 2016; 133, 273-281.
- Jacob S, Shahzad MA, Maheshwari R, Panaich SS, Aravindhakshan R. Cardiac rhythm device identification algorithm using X-Rays: CaRDIA-X. Heart Rhythm 2011; 8, 915-922.
- Nakou ES, Simantirakis EN, Kallergis EM, Nakos KS, Vardas PE. Cardiac resynchronization therapy (CRT) device replacement considerations: upgrade or downgrade? A complex decision in the current clinical setting. Europace 2017; 19, 705-711.
Version History: This Fast Fact was originally edited by David E Weissman MD and published in April 2004. Re-copy-edited in April 2009; then again in June 2015 – reference 5 added and incorporated into text. In November 2017 it underwent further revision to add a discussion regarding cardiac resynchronization therapy devices.
Fast Facts and Concepts are edited by Sean Marks MD (Medical College of Wisconsin) and associate editor Drew A Rosielle MD (University of Minnesota Medical School), with the generous support of a volunteer peer-review editorial board, and are made available online by the Palliative Care Network of Wisconsin (PCNOW); the authors of each individual Fast Fact are solely responsible for that Fast Fact’s content. The full set of Fast Facts are available at Palliative Care Network of Wisconsin with contact information, and how to reference Fast Facts.
Copyright: All Fast Facts and Concepts are published under a Creative Commons Attribution-NonCommercial 4.0 International Copyright (http://creativecommons.org/licenses/by-nc/4.0/). Fast Facts can only be copied and distributed for non-commercial, educational purposes. If you adapt or distribute a Fast Fact, let us know!
Disclaimer: Fast Facts and Concepts provide educational information for health care professionals. This information is not medical advice. Fast Facts are not continually updated, and new safety information may emerge after a Fast Fact is published. Health care providers should always exercise their own independent clinical judgment and consult other relevant and up-to-date experts and resources. Some Fast Facts cite the use of a product in a dosage, for an indication, or in a manner other than that recommended in the product labeling. Accordingly, the official prescribing information should be consulted before any such product is used.